| Model Number |
83030-01 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on 10/14/2024, the chipboard box of the xact self expanding stent system (sess) was noted to have the catheter length labeled as "60 mm" rather than "60 cm".The catheter length is labeled on numerous areas of device, including the chipboard box, pouch label and ifu.All other locations where the catheter length is labeled were correct in indicating 60cm.The stent was implanted on (b)(6) 2024 without issue in the left internal carotid artery with heavy calcification, type 3 arch and moderate tortuosity.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Electronic lot history record (elhr) and corrective and preventive actions (capa) reviews were performed and revealed an indication of a product quality issue.Additionally, a review of the complaint handling database identified one other similar incidents from this lot.The investigation determined the reported device markings / labelling problem appears to be a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.E1: (b)(6).
|
| |
|
Search Alerts/Recalls
|