SMITH & NEPHEW, INC. DYONICS POWERMINI, WITH HAND CONTROLS; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72201500
Medical Device Problem Codes Mechanical Problem (1384); Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a central reprocessing, the dyonics powermini overheated.No burn or damage occurred to the user.There was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.During testing, a functional evaluation revealed device grew warm and a blade stall error message was found.It was also noted device ma was above 960.Further investigation revealed stiffness in the drive fork when rotated.These failures have been determined to be related to the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with a mechanical component failure.Factors that could have contributed to the failure include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS POWERMINI, WITH HAND CONTROLS
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20627497
Report Number1643264-2024-00596
Device Sequence Number15658676
Product Code HAB
UDI-Device Identifier03596010621696
UDI-Public3596010621696
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2024
Supplement Date Received by Manufacturer11/27/2024
Initial Report FDA Received Date11/07/2024
Supplement Report FDA Received Date11/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-