Zimvie complaint number (b)(4).D10: concomitant medical product: spwb10, impl tapered sp 4.8mm 10m m octagon, lot: 2019010023 spwb10, impl tapered sp 4.8mm 10m m octagon, lot: 2019010023 spwb10, impl tapered sp 4.8mm 10m m octagon, lot: 2019010134 g4: premarket identification: k011245|k002188 h6: event problem codes: patient code: 1932.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|