| Model Number |
DVPC3D4 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
Dizziness (2194); Shock from Patient Lead(s) (3162)
|
| Date of Event |
10/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced dizziness.The health care professional (hcp) stated the device shocked the patient as they were being brought to the hospital.There was no arrythmia noted since the last interrogation.No additional information could be obtained through follow-up.The implantable cardioverter defibrillator (icd) remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|