• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KEYSTONE DENTAL, INC. PRIMA; PRIMACONNEX TAPERED WD 5.0X13

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KEYSTONE DENTAL, INC. PRIMA; PRIMACONNEX TAPERED WD 5.0X13 Back to Search Results
Model Number 15423K
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Osteopenia/ Osteoporosis (2651)
Date of Event 08/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Keystone dental inc.Provides labeling with all implants.Failure to osseointegrate and loss of osseointegration (implant mobility) is identified as an adverse reaction and identified in the list of warnings in all endosseous dental implant labeling.The specific cause for a particular complaint is often not readily identified as there are various factors that contribute to the risk of implant failure.These include patient factors such as prior oral infection, poor bone quality or quantity, systemic conditions such as diabetes and uncontrolled hypertension.Technique factors such as overheating of the implant site may cause necrosis of the bone, preventing growth (pelayo et al., 2008; turkyilmaz & al., 2008).Patient habits such as tobacco use (heitz-mayfield & huynh-ba, 2009), alcohol or drug abuse, poor oral hygiene, and bruxism (salvi & braegger, 2009) may also lead to implant failure.In addition, improper surgical technique can lead to implant failure and/or loss of supporting bone (salvi & braegger, 2009).
 
Event or Problem Description
Failed to osseointegrate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMA
Common Device Name
PRIMACONNEX TAPERED WD 5.0X13
Manufacturer (Section D)
KEYSTONE DENTAL, INC.
154 middlesex turnpike
burlington MA 01803
Manufacturer (Section G)
KEYSTONE DENTAL, INC.
154 middlesex turnpike
burlington MA 01803
Manufacturer Contact
erik roman
154 middlesex turnpike
burlington, MA 01803
7813283490
MDR Report Key20631039
Report Number3005990499-2024-72135
Device Sequence Number7929405
Product Code DZE
UDI-Device IdentifierD76815423K0
UDI-PublicD76815423K0
Combination Product (Y/N)N
Initial Reporter CountryTW
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 11/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number15423K
Device Catalogue Number15423K
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2024
Initial Report FDA Received Date11/07/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-