• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SMARTSET HV BONE CEMENT 40G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SMARTSET HV BONE CEMENT 40G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3092040
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: e1: reporter is a legal professional.
 
Event or Problem Description
On (b)(6) 2016, patient received an attune right total knee arthroplasty to address primary osteoarthritis.They received a size 7 posterior stabilized femur, a size 6 tibial tray, a 6mm thick insert, and 38mm patella, all using depuy smartset hv cement.There were no complications.On (b)(6) 2020, patient's right knee was revised to address aseptic implant loosening of tibia and femur.Intraoperatively, the tibial baseplate was found to be loose at the cement to implant interface; surgeon reported that there was no cement adhered/bonded to the backside of the tibial tray at all.The femur was also loose and explanted as well; there was cement bonded to the femur component.The femur loosening interface was both cement to implant and cement to bone loosening.The patella component was well-fixed and retained.Surgeon re-implanted competitor products, with no reported complications.Doi: (b)(6) 2016.Dor: (b)(6) 2020.Right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET HV BONE CEMENT 40G
Common Device Name
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20631264
Report Number1818910-2024-23262
Device Sequence Number7909660
Product Code LOD
UDI-Device Identifier10603295168331
UDI-Public(01)10603295168331(17)20170630(10)8143804
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K033563
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2017
Device Catalogue Number3092040
Device Lot Number8143804
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/29/2024
Supplement Date Received by Manufacturer11/18/2024
Initial Report FDA Received Date11/07/2024
Supplement Report FDA Received Date11/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/06/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE MEDIAL DOME PAT 38MM; ATTUNE PS FB INSRT SZ 7 6MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
-
-