| Model Number |
1000186-115 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.On oct 24th 2024, abbott vascular determined that a field action was required for specific lots of priority pack 20/30 w/115 rhv product.The product associated with this complaint is from the impacted population.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure, the 20/30 priority pack indeflator with rotating hemostatic valve was connected directly to an unspecified balloon when a leak was noted.Air was noted to enter through the sealing of the hose, as seen by air bubbles, and the pressure on the gauge decreased.Another indeflator was used to continue the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported leak was unable to be confirmed.Device history record (dhr) and corrective and preventive actions (capa) reviews were performed and revealed an indication of a product quality issue.Additionally, a review of the complaint handling database identified similar incidents from this lot.The investigation determined the reported difficulties appear to be a potential product quality issue.On october 24th, 2024, abbott vascular determined that a field safety corrective action is required for specific lots of 20/30 priority pack with copliot, 20/30 indeflator, priority pack 20/30 w/115 rhv, and ppak 20/30 with rhv.This action is being taken due to an increased potential for a leak in the indeflator under pressure or vacuum due to a gap in the hose snap fitting and o-rings from a specific supplier.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|