| Catalog Number |
ECR45W |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a laparoscopic lobectomy procedure, could not fire farther after fired a little.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 11/7/24.D4: batch #unk.Additional information was requested, and the following was obtained: - did the device lock-out (no staples deployed and no cut line started)? or, did the device start to deploy staples and cut but could not be completed (partially fire)? or, did the device fully fire and stop during the automatic knife return? partially fired was there any difficulty opening the device jaws? no.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device ecr45w, lot number 137c01; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 11/21/2024 the device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 12/3/2024.D4 batch #917a36.Investigation summary: the product was returned or evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that one ecr45w reload was received unfired and with one tyvek.The returned reload was loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.The event described could not be confirmed as the reload was performed without any difficulties noted.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: insert the reload by sliding it against the bottom of the cartridge jaw until the cartridge alignment tab stops in the reload alignment slot.Snap the reload securely in place.Remove the staple retaining cap and discard.Refer to the echelon endoscopic linear cutter insert for instructions for using the instrument after it is loaded.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 917a36, and no non-conformances were identified.
|
| |
|