MAUDE Adverse Event Report: DATASCOPE CORP. MEGA 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. MEGA 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0498-01
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Udi # is incomplete as the lot #, manufacture date, exp.Date were not provided.This information was requested from the customer; however, despite our best efforts, no information has been received.If any pertinent information is received in the future, a supplemental report will be submitted.The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event or Problem Description
During the placement of the intra-aortic balloon (iab), the guidewire was bent.The guidewire from another iab was used, which also bent.A spare guidewire was then used, which allowed the balloon to be placed.This report is for the 2nd product used in this event.A separate report will be submitted for the 1stproduct used.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint # (b)(4).
 
Additional Manufacturer Narrative
Corrected fields: d4- unique identifier (udi)# no decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Complaint record id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA 8FR. 50CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer Contact
henry pena
15 law drive
fairfield, NJ 07004
MDR Report Key20635035
Report Number2248146-2024-0000681
Device Sequence Number12634661
Product Code DSP
UDI-Device Identifier10607567108018
UDI-Public(01)10607567108018(10)3000308474(17)230502(11)260502
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/02/2026
Device Model Number0684-00-0498-01
Device Catalogue Number0684-00-0498-01
Device Lot Number3000308474
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2024
Supplement Date Received by Manufacturer12/05/2024
02/04/2025
Initial Report FDA Received Date11/07/2024
Supplement Report FDA Received Date12/05/2024
02/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient Age56 YR
Patient SexMale
Patient Weight94 KG
-
-