| Model Number |
0684-00-0498-01 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Udi # is incomplete as the lot #, manufacture date, exp.Date were not provided.This information was requested from the customer; however, despite our best efforts, no information has been received.If any pertinent information is received in the future, a supplemental report will be submitted.The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
During the placement of the intra-aortic balloon (iab), the guidewire was bent.The guidewire from another iab was used, which also bent.A spare guidewire was then used, which allowed the balloon to be placed.This report is for the 2nd product used in this event.A separate report will be submitted for the 1stproduct used.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected fields: d4- unique identifier (udi)# no decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Complaint record id # (b)(4).
|
| |
|