| Model Number |
Z800WF |
| Medical Device Problem Codes |
Excess Flow or Over-Infusion (1311); Failure to Power Up (1476); Calibration Problem (2890); Infusion or Flow Problem (2964); Power Problem (3010); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There are no previous complaints on the device related to this issue.This pump underwent routine maintenance testing where it was detected the pump's power test failed and it's performance felled outside the acceptable range for the flowrate %.Consequently, it necessitated to replace the power componet and adjustment to the pump's flowrate %.It was confirmed the replacing of the componet and recalibration addressed the issue successfully.A capa has been opened in order to fully investigate and address the root cause of the reported event.Reference to complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
On 10/23/2024, znyo medical received a report that during the preventive maintenance (pm), the device power on failed and there was a flow rate deviation at 11.83%.No report of patient injured/harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment, the reported flow rate failure mode has been determined to be non-reportable.Events involving a faster flow rate are not reportable when flow rate accuracy is above +5% but below +15% specification.The severity of this failure is 1 - inconvenience or temporary discomfort.Our evaluation concluded that the event did not pose a risk to the health of patients, users, or bystanders.Furthermore, the report did not allege a serious injury or death, nor would a recurrence of the reported issue be expected to cause or contribute to a serious injury or death.Reference to complaint #: (b)(4).
|
| |
|