• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 1708070
Medical Device Problem Codes Expulsion (2933); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Skin Erosion (2075)
Date of Event 12/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through litigation process that sometime post a port placement, the patient allegedly developed infection.However, the current status of the patient is unknown.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation, medical record was provided for review.Medical record alleges that, a patient was implanted with power port implantable port via right cephalic vein for multiple complicated medical issues, including autonomic dysfunction, paroxysmal atrial fibrillation, ehlers danlos syndrome, and postural orthostatic tachycardia syndrome and frequently needs iv fluid infusions for postural orthostatic tachycardia syndrome (pots).Intraoperative fluoroscopy confirmed the catheter tip at the cavo atrial junction, with good blood return and saline flush.Approximately, two years and ten months later, the patient presented with complaints of port eroding and chronic pain and discussed for removal and replacement of a port.Subsequently, on the same day port was removed and new port was implanted.The investigation is inconclusive for the reported expulsion as no evidence was provided in the provided report.Furthermore, the clinical conditions alleged in the complaint can be confirmed.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B3, b5, g3, h6 (patient, device, method) section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the litigation process that, on (b)(6) 2020, a patient underwent port placement via the right cephalic vein for intravenous access related to postural orthostatic tachycardia syndrome (pots).Approximately two years, nine months, and twenty six days later, on (b)(6) 2022, the patient presented with port erosion and chronic pain.It was further reported that the patient was diagnosed with an unspecified infection.Subsequently, the port was removed on the same day, and a new port was implanted.However, the current status of the patient remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key20638077
Report Number3006260740-2024-06534
Device Sequence Number7385748
Product Code LJT
UDI-Device Identifier00801741026652
UDI-Public(01)00801741026652(17)200831(10)RECX2357
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K060812
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Catalogue Number1708070
Device Lot NumberRECX2357
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2024
Supplement Date Received by Manufacturer04/29/2026
Initial Report FDA Received Date11/08/2024
Supplement Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexFemale
-
-