RESPIRONICS, INC. DREAMSTATION AUTO BIPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number AHX700H11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Pulmonary Embolism (1498); Cancer (3262); Unspecified Respiratory Problem (4464)
Date of Event 10/09/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information that the patient alleging bladder cancer, nodule on left lung and pulmonary embolism.There was no report of medical intervention.The manufacturer was made aware of this complaint through a representative of the customer.The device was returned to the third-party service centre for further evaluation.During the evaluation of the device at the third-party service centre, the device was visually inspected and found no evidence of foam degradation.During the evaluation, secondary findings observed.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service centre.There was no error code found.The third-party service centre concludes that they couldn't confirm the customer's allegation and there was no visible foam degradation and unit got scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO BIPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key20638384
Report Number2518422-2024-69097
Device Sequence Number7471111
Product Code BZD
UDI-Device Identifier00606959022638
UDI-Public(01)00606959022638
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 11/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberAHX700H11
Device Catalogue NumberAHX700H11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2023
Initial Report FDA Received Date11/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/01/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-