| Model Number |
CF-HQ1100DI |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, the jet tube of the colonovideoscope was clogged with foreign material.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.It is unknown if there was deviation from the instructions for use in the reprocessing steps for the location where the foreign material remained.Additionally, there was no physical damage on the location where the foreign material remained.The event can be detected/prevented by following the instructions for use which state: detection method is described in cf-hq1100dl/i operation manual chapter 3 preparation and inspection.Preventive measures are described in cf-hq1100dl/i reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|
| |
|