• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION SABERX RADIANZ; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SABERX RADIANZ; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number R51008004L
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the 8mm 4cm saberx radianz percutaneous transluminal angioplasty (pta) balloon ruptured at 8atm while being used during an evt case.It was replaced with a new saber pta (radianz) (8*6), and the procedure was completed.There was no reported injury to the patient.The product was stored properly according to the instructions for use (ifu).There were no difficulties in removing the product from the hoop, the protective balloon cover, the stylet, or any sterile packaging components.No kinks or damages were noted prior to inserting the product into the patient.The device was prepped per the instructions for use (ifu), maintaining negative pressure during the preparation.The intended procedure targeted the iliac region, with the lesion characterized as having no calcification, no vessel tortuosity, and 90% stenosis.The device was not used for a chronic total occlusion (cto) there was no resistance or friction while inserting the balloon through the rotating hemostatic valve or the guide catheter.Additionally, there were no difficulties advancing the balloon catheter through the vessel, crossing the lesion with the balloon catheter.The catheter was never in an acute bend, and the balloon catheter did not kink while being used.The device was discarded.
 
Additional Manufacturer Narrative
As reported, the 8mm 4cm saberx radianz percutaneous transluminal angioplasty (pta) balloon ruptured at 8atm while being used during an evt case.It was replaced with a new saber pta (radianz) (8*6), and the procedure was completed.There was no reported injury to the patient.The product was stored properly according to the instructions for use (ifu).There were no difficulties in removing the product from the hoop, the protective balloon cover, the stylet, or any sterile packaging components.No kinks or damages were noted prior to inserting the product into the patient.The device was prepped per the instructions for use (ifu), maintaining negative pressure during the preparation.The intended procedure targeted the iliac region, with the lesion characterized as having no calcification, no vessel tortuosity, and 90% stenosis.The device was not used for a chronic total occlusion (cto) there was no resistance or friction while inserting the balloon through the rotating hemostatic valve or the guide catheter.Additionally, there were no difficulties advancing the balloon catheter through the vessel, crossing the lesion with the balloon catheter.The catheter was never in an acute bend, and the balloon catheter did not kink while being used.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.The reported severe stenosis of the iliac vessel may have contributed to the reported event; it is likely to have caused damage to the balloon material that led to its rupture.However, without return of the product for analysis or films of the event it is difficult to draw a clinical conclusion between the device and the reported event.According to the warnings in the safety information in the instructions for use, ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ the information available does not suggest a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABERX RADIANZ
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
coyol free zone, building b25, ,  , costa rica
el coyol alajuela bld 25 CR201-02
CS   CR20102
7863138372
MDR Report Key20639511
Report Number3007635982-2024-00266
Device Sequence Number17997801
Product Code LIT
UDI-Device Identifier20705032081471
UDI-Public(01)20705032081471(17)270228(10)82307514
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201377
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberR51008004L
Device Lot Number82307514
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SABER PTA (RADIANZ) (8*6).; SHEATH (TERUMO).
Patient SexUnknown
-
-