| Model Number |
799-EGL2-00 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was evaluated by zoll medical corporation.The customer's report was duplicated and attributed to capacitor on the smart pneumatic module board.The smart pneumatic module board will be replaced to resolve the report.The device will be recertified and returned to the customer once the repair estimate has been approved.Analysis for reports of this type has not identified an increase in trend.
|
| |
|