MAUDE Adverse Event Report: COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE Back to Search Results
Catalog Number ES2816
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2024
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information, the implant was removed and replaced due to leak.
 
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
 
Additional Manufacturer Narrative
Titan touch pump, cylinders 1 and 2, reservoir, and detached exhaust tubing were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.Abrasion was noted on the exhaust tubing of cylinder 1.The surfaces of the detachment site of the exhaust tube of cylinder 1 appear to be rough and irregular, indicating stress was exerted.No functional abnormalities were noted with cylinder 2.No functional abnormalities were noted with the reservoir.Abrasion was noted on the piece of detached exhaust tubing.The surfaces of the detachment site of the detached exhaust tubing appear to be rough and irregular, indicating stress was exerted.Based on examination of the returned product, it was concluded that while in-vivo both the exhaust tubes and inlet tube of the pump had overlapped and abraded against one another.While no functional abnormalities were noted during testing with the returned components, microscopic examination of the surfaces of the exhaust tube detachment end between the piece of detached exhaust tubing and cylinder 1 revealed rough and irregular surfaces, indicating stress was exerted.Based on the overlapping of the pump tubes, in combination with device usage over time, this could contribute to sufficient stress to separate the exhaust tubing of cylinder 1.A separation of this type would then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Event or Problem Description
According to the available information, the implant was removed and replaced due to leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key20640249
Report Number2125050-2024-01826
Device Sequence Number15309227
Product Code FHW
UDI-Device Identifier05708932539111
UDI-Public05708932539111
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/12/2021
Device Catalogue NumberES2816
Device Lot Number4949402_ES28162400
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/08/2024
Supplement Report FDA Received Date01/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexUnknown
-
-