| Catalog Number |
8065752918 |
| Medical Device Problem Codes |
Suction Problem (2170); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Other health care professional reported that pressure was unstable during phaco mode and abnormal sound heard at fluidic management system (fms).The procedure type was unknown.No patient impact was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information was updated in d.4 the manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in d.9., h.3., h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A used active fluidic management system in a tray were visually inspected and was tested using a console.The sample could be recognized by the console.The sample primed and tuned with a handpiece and a 0.9millimiter aspiration bypass system (abs) tip and infusion sleeve from lab stock successfully and the service data could be retrieved.No system message code displayed on console screen.The sample functioned within specification.Cassette maximum vacuum and aspiration flow rate, and irrigation flow rate were measured and met specifications.No problem was found with the returned cassette fluidics.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.After an investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|