Zimvie complaint (b)(4).Multiple reports are being submitted for this event.A4: patient weight is not provided / unknown.E1: initial reporter¿s title is not provided / unknown.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformance's affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
|
It was reported that after uncovering, implant threads became visible.Atrophy around all implants.Tooth #'s 7, 10, 13, and 27 (universal).The implants were removed due to bone loss.
|