| Model Number |
2R8546 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a clearlink system continu-flo solution set leaked at the y-site port despite having a green alcohol cap in place.This was observed during daily audits.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: unique identifier (udi) #: the lot number (10) and subsequent expiration date (17) are unknown; therefore, the udi number only will contain the device identifier (01).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the actual device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Pressure and clear passage testing, priming, usage testing by gravity, usage testing using an in-lab spectrum pump, and water referee back pressure testing on the clearlinks were performed with no issues observed.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|