| Model Number |
BF-1T150 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a retrospective complaint review involving bf series scopes.This report reflects olympus' re-evaluation of previous medical device reporting decision.The device was returned for evaluation where the reported event was identified.Based on the results of the investigation, a definitive root cause cannot be identified.The probable cause was the use of high-energy hand instruments (laser/high-frequency/etc) without ensuring sufficient distance from the distal end (handling problem).A device history review revealed no issues that could have caused or contributed to the reported issue.The suggested event can be detected and prevented by handling the device in accordance with ¿chapter 4 operation¿ in ifu.Olympus will continue to monitor the field performance of this device.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the bronchovideoscope¿s plastic distal end cover was burned or melted around the instrument channel outlet at the distal end.There were no reports of patient involvement.
|
| |
|
Search Alerts/Recalls
|