| Model Number |
1000186-115 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during preparation the indeflator was noted to have a leak as the device was not able to hold negative when connected directly to a balloon prior to being used in the procedure.There was no patient involvement and no clinically significant delay reported in the procedure.Another indeflator was used in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
On oct 24th 2024, abbott vascular determined that a field action was required for specific lots of priority pack 20/30 w/115 rhv product.The product associated with this complaint is from the impacted population.Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported leak was unable to be confirmed.Device history record (dhr) and corrective and preventive actions (capa) reviews were performed and revealed an indication of a product quality issue.Additionally, a review of the complaint handling database identified similar incidents from this lot.The investigation determined the reported difficulties appear to be a potential product quality issue.On oct 24th 2024, abbott vascular determined that a field action was required for specific lots of priority pack 20/30 w/115 rhv product.The product associated with this complaint is from the impacted population.This action is being taken due to an increased potential for a leak in the indeflator under pressure or vacuum due to a gap in the hose snap fitting and o-rings from a specific supplier.
|
| |
|
Search Alerts/Recalls
|