| Model Number |
5016000 |
| Medical Device Problem Codes |
High Readings (2459); Low Readings (2460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, prior to use, the device's higher temperature accuracy was off.The high temperature was 49c instead of 47c, and the low temperature was 35.6c.The customer followed the service manual to perform the temp test on the unit and had tested 10 other units that were in spec.The customer used a fluke meter with a type k thermocouple.There was no patient involvement.
|
| |
|