It was learned through implant patient registry and investigation that a 23mm 3000 aortic valve in pulmonic position was disabled via valve in valve procedure after an implant duration of eight (8) years, ten (10) months due to stenosis and regurgitation.The patient presented with shortness of breath, fatigue, and activity intolerance.Tpvr was completed with a 23mm 9750tfx transcatheter valve and patient was stable at the end of the procedure.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
|
|
H11: additional manufacturer narrative: updated sections: b4, g3, g6, h2, h6 (type of investigation, investigation findings, and investigation conclusions).The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: based on the information available, the device was intentionally used outside the way it was designed and intended.The performance, safety, efficacy, longevity, and durability of rings and bioprosthetic valves have not been determined when the product is used outside the scope of its intended use and/or its studied population.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Per technical summary 33069, rev a, structural valve deterioration (svd) is the most common reason for bio prosthesis explant and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed, and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A dhr review was performed, and no related manufacturing nonconformances were identified.There is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.Per (b)(4), rev q, and (b)(4), rev o, a capa/scar/pra is not required, as there are no confirmed product or labeling non-conformances, and no other triggers are met.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.All pertinent information available to edwards lifesciences has been submitted.
|