| Model Number |
ESBF2514C103E |
| Medical Device Problem Codes |
Leak/Splash (1354); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cardiac Arrest (1762); Rupture (2208)
|
| Date of Event |
12/18/2022
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
An endurant iis stent graft was implanted during the endovascular treatment of a ruptured 45mm aaa.It was reported that the patient had a type i endoleak at the end of the procedure.It was reported approximately 2 years post the index procedure intervention was performed for a rupture due to the type ia.10 end oanchors were placed, however on the post-angio the type ia endoleak remained.Per the physician the cause of the endoleak was due to under-sizing.No additional clinical sequalae were provided and the patient will be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received : it was reported that the main body and limbs were explanted.As the patient was moved to the hospital bed, the patient had a heart attack and expired in the or.Per the physician, the cause of death was valvular heart disease.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|