| Model Number |
N/A |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Clinical code: 4582 - no clinical signs, symptoms or conditions: no patient consequences were identified during initial event intake form.Health impact: 2199 - no health consequences or impact: no patient consequences were identified during initial event intake form.Medical device problem: 1494 - off- label use: the device is contraindicated for ex anantomical use.Component code: 4755 - part/component/sub-assembly term not applicable.Type of investigation: 4110 - trend analysis: 4 year review was performed for leakage > event description (circulation line) with gave an occurrence rate of (b)(4).Increasing trend was identified requiring action at this time.4111 - communication/interview: awaiting further information from the site.3331 - analysis of production records: a comprehensive batch review was performed to review the manufacturing process from raw materials to finished product.No issues were identified.No causal link was identified between the device and the event.4117 - device not accessible for testing: device not returned by the site for testing.Investigation findings: 3233 - results pending completion of investigation: investigation conclusion: 11 - conclusion not yet available:.
|
| |
|
Event or Problem Description
|
|
Event date: on 16 oct 2024.Implant date: on (b)(6) 2024.Event was reported as weaving defects.The surgical graft was used as a circulation line for aortic replacement procedure, and after clamp released, surgeon found blood eject from the surface of surgical graft (not the anastomosis site).
|
| |
|
Additional Manufacturer Narrative
|
|
Clinical code: 4582 - no clinical signs, symptoms or conditions: no patient consequences were identified during initial event intake form.Health impact: 2199 - no health consequences or impact: no patient consequences were identified during initial event intake form.Medical device problem: 1494 - off- label use: ifu 301-199 polyester multi-product row gelita set b states in section 1.3 contraindications.Gelsoft and gelsoft plus prosthesis are contraindicated for thoracic and extra anatomical use.Component code: 4755 - part/component/sub-assembly term not applicable.Type of investigation: 4110 - trend analysis: 4 year review was performed for leakage > event description (circulation line) with gave an occurrence rate of 0.002%.Increasing trend was identified requiring action at this time.4111 - communication/interview: additional information was received by the site which confirmed that: the size of the hole was 0.2mm-0.5mm and the hole was noticed when there was perfusion to the graft.The hole was fixed with a suture.Atraumatic clamp were used.The patient total blood loss was 10-15ml, the graft was used for bypass.The surgeon decided to use a gelsoft plus device instead of gelweave device for circulation line.There was no damage to graft before use and no further device was used.The patient was fine and without adverse event.3331 - analysis of production records: a comprehensive batch review was performed to review the manufacturing process from raw materials to finished product.No issues were identified.No causal link was identified between the device and the event.4117 - device not accessible for testing: device not returned by the site for testing.Investigation findings: 4248 - usage problem identified: ifu 301-199 polyester multi-product row gelita set b states in section 1.3 contraindications.Gelsoft and gelsoft plus prosthesis are contraindicated for thoracic and extra anatomical use.Investigation conclusion: 18 - failure to follow instructions: ifu 301-199 polyester multi-product row gelita set b states in section 1.3 contraindications.Gelsoft and gelsoft plus prosthesis are contraindicated for thoracic and extra anatomical use.The root cause of this event was determined to be off-label use.24 - cause traced to intentional off-label, unapproved, or contraindicated use: ifu 301-199 polyester multi-product row gelita set b states in section 1.3 contraindications.Gelsoft and gelsoft plus prosthesis are contraindicated for thoracic and extra anatomical use.
|
| |
|
Event or Problem Description
|
|
Event date: 16 oct 2024, implant date: on (b)(6) 2024.Event was reported as weaving defects.The surgical graft was used as a circulation line for aortic replacement procedure, and after clamp released, surgeon found blood eject from the surface of surgical graft (not the anastomosis site).This report is being submitted as a follow up 1 for manufacturing report#: 9612515-2024-00077 to provide event closure information for tag0065.
|
| |
|