ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 15X150MM; PROTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection of stock, the sterile package was found damaged.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
(b)(4).G2: foreign: country: japan the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister); however, sterility had not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 15X150MM
Common Device Name
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20650945
Report Number0001825034-2024-02634
Device Sequence Number12422386
Product Code KWA
UDI-Device Identifier00880304491861
UDI-Public(01)00880304491861(17)301205(10)6893575
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101086
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-103150
Device Lot Number6893575
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2024
Supplement Date Received by Manufacturer03/07/2025
Initial Report FDA Received Date11/11/2024
Supplement Report FDA Received Date03/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-