MEDTRONIC, INC. TEMPORARY EXTERNAL PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 53401
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the external pulse generator (epg) powered off unexpectedly a few minutes after powering on.The epg was checked several times however the issue persisted after changing the epg batteries.There was no patient involvement.
 
Additional Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment that the external pulse generator (epg) powered off unexpectedly.The epg battery contact was corroded.All found defective parts were replaced and all other identified issues were resolved.The epg then passed all final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPORARY EXTERNAL PACEMAKER
Common Device Name
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key20651303
Report Number2182208-2024-04787
Device Sequence Number8283968
Product Code DTE
UDI-Device Identifier00643169723160
UDI-Public00643169723160
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K162054
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number53401
Device Catalogue Number53401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2024
Initial Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-