• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX BLUSELECT PLAIN INNER CANNULA; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX BLUSELECT PLAIN INNER CANNULA; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 101/858/080
Medical Device Problem Code Misconnection (1399)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.Date of event: month and year of event have been provided, but day is unknown.H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the inner trach cannula did not positively click into place when performing a trach change on a patient in the intensive care unit (icu).The rib around the inner cannula wasn't clicking into the outer cannula, it was sliding in and out without interference, leading to possible ¿popping out¿ of the inner cannula.The health professional tried a second one to find the same issue.There was patient involvement, and no patient harm/adverse event reported.
 
Additional Manufacturer Narrative
D4 - expiration date and d4 - primary udi number; corrected.H6.Codes: updated.No product or photos were returned therefore no device analysis could be completed.A device history record (dhr) review showed there were no discrepancies or non-conformances during the manufacturing of the reported lot number.If the product is returned the manufacturer will re-open the complaint for further device analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUSELECT PLAIN INNER CANNULA
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key20652108
Report Number9617604-2024-00859
Device Sequence Number8219094
Product Code BTO
UDI-Device Identifier15019517077865
UDI-Public(01)15019517077865(17)281114(10)4442599
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number101/858/080
Device Lot Number4442599
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2024
Supplement Date Received by Manufacturer11/11/2024
Initial Report FDA Received Date11/11/2024
Supplement Report FDA Received Date12/04/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-