MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFT015
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Codes Visual Disturbances (2140); Insufficient Information (4580)
Date of Event 09/17/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for analysis.Complaint history and product history records could not be reviewed because the reporter did not provide a lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A surgeon reported that following an intraocular lens (iol) implant procedure, observed scratch on the lens.The iol was exchanged for company lens following the initial implant procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information was provided in h.11.The product was not returned.Complaint history and product history records could not be reviewed because the reporter did not provide a lot number or any identification traceable to the manufacturing documentation.Associated products were not provided.It is unknown if qualified associated products were used.The product investigation could not identify a root cause for the reported complaint.The product was not returned.Not enough information was provided for further investigation.It is unknown if qualified associated products were used.The instructions for use (ifu instructs: company foldable intraocular lens (iols) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Due diligence has been performed in an attempt to obtain further information on this event.The facility has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned.Complaint and product history records were reviewed, and documentation indicated the product met release criteria.A qualified cartridge and handpiece were indicated with a non-qualified viscoelastic.The root cause for the reported scratch may be related to a failure to follow the instructions for use (ifu).A non-qualified viscoelastic was indicated.The ifu instructs: company foldable intraocular lens (iols) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The product investigation could not identify a root cause for the photophobia.Information was provided that the company, 24.0 diopter lens was replaced with a company, 22.5 diopter lens.The 1.50 changed in diopter may indicate the replacement lens was an improved choice for the patient's visual needs.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
Additional information has been requested and received stating that following an intraocular lens (iol) implant procedure, the patient experienced photophobia, light scattering.Clinical reason for explant was scratched iol.The iol was exchanged for same model different diopter advanced technology intraocular lenses (atiol) company lens 4 weeks following the initial implant procedure.In the surgeon opinion iol against cartridge system had caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20653578
Report Number1119421-2024-02187
Device Sequence Number15524781
Product Code POE
UDI-Device Identifier00380652421317
UDI-Public(01)00380652421317(17)260211(11)210212(10)15148395
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2026
Device Model NumberDFT015
Device Catalogue NumberDFT015.240
Device Lot Number15148395
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/16/2024
Supplement Date Received by Manufacturer01/14/2025
04/21/2025
Initial Report FDA Received Date11/11/2024
Supplement Report FDA Received Date02/09/2025
05/15/2025
Date Device Manufactured02/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEALON PRO; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
-
-