• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-Q190
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a retrospective complaint review involving bf series scopes.This report reflects olympus' re-evaluation of previous medical device reporting decision.The device was returned for evaluation where the reported event was identified.Based on the results of the investigation, a definitive root cause cannot be identified.The probable cause was the use of high-energy hand instruments (laser/high-frequency/etc) without ensuring sufficient distance from the distal end (handling problem).A device history review revealed no issues that could have caused or contributed to the reported issue.Labeling: this issue is addressed in the instructions for use (ifu): the suggested event can be detected and prevented by handling the device in accordance with ¿chapter 4 operation¿ in ifu.Olympus will continue to monitor the field performance of this device.
 
Event or Problem Description
It was observed that during device evaluation, the bronchovideoscope's plastic distal end cover was burned/melted around the instrument channel outlet at the distal end.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20653915
Report Number9610595-2024-23151
Device Sequence Number12878024
Product Code EOQ
UDI-Device Identifier04953170434792
UDI-Public04953170434792
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K121959
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-Q190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Initial Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2011-2023 AND Z-0189-2
Patient Sequence Number1
Patient SexUnknown
-
-