This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added to the following field: h3, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, malfunction most likely occurred by physical stress due to bumping or dropping the distal end of the scope, or chemical stress due to the chemical solution used.However, a root cause could not be determined.The events can be detected and prevented in accordance with the following sections of the instructions for use: "operation manual for ltf-s190-10 important information: please read before use: dangers, warnings, and cautions: do not strike, hit, or drop the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Reprocessing manual for ltf-s190-10 3.1 compatibility summary methods listed as ¿compatible¿ in table 3.1 and table 3.2 are compatible for routine use only when used according to manufacturers instructions.Repeated use and reprocessing of endoscopes and accessories lead to gradual wear and tear.Furthermore, reprocessing methods that employ higher temperatures and more caustic/corrosive materials may lead to faster deterioration.In general, sterilization processes are harsher on equipment than disinfection processes.Before each patient procedure, inspect the endoscope and accessories for damage, according to the instructions described in this manual and its companion ¿operation manual¿.3.1 compatibility summary: list of compatible methods validated in terms of biological efficacy and material durability sterilization by sterrad® 50/100s/200/nx¿ system may deteriorate the adhesive of the insertion section.Depending on the circumstances, replacement of the insertion section may be required.Before use, confirm that the endoscope is free from any damage or other irregularities.For further details, contact olympus." olympus will continue to monitor field performance for this device.
|