This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Pump was received with a critical pump error (open book image) alarm.Unable to perform the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement test, active current measurement test and self test due to critical pump error (open book image) alarm.Pump was cut open to perform visual inspection and found moisture damage on the force sensor.Force sensor zero offset within specification (22.76 mv).Test p-cap and reservoir locked properly into reservoir compartment during testing.The following were noted during visual inspection: faded serial number label, cracked battery tube threads, cracked case at corner of the belt clip rails, cracked case along the side of the battery tube and cracked retainer ring.Critical pump error (open book image) alarm confirmed due to moisture damage on the force sensor.Cosmetic damage confirmed at the battery compartment.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|