| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Head Injury (1879); Failure of Implant (1924)
|
| Date of Event |
09/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The recipient's hearing performance with the device is affected.Further technical checks are considered.However, med-el clinical support recommended re-implantation.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
| |
|
Event or Problem Description
|
|
The recipient's hearing performance with the device is affected.Re-implantation was done on (b)(6) 2025.
|
| |
|
Event or Problem Description
|
|
The recipient's hearing performance with the device is affected.Re-implantation is scheduled for (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress caused by the reported trauma appears very likely.However, to determine an exact root cause device investigation would be necessary.Reimplantation is scheduled.
|
| |
|