| Catalog Number |
136528720 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
10/28/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.E1 initial reporter address line 1: (b)(6).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent left hip revision due to pain.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: révision hip left.The product was not returned to j&j medtech orthopaedics, however photos were provided for review.See attachment image.Jpg, image.Jpg.The photographs attached were reviewed, however they do not represent the reported complaint.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the photographs provided are not representative of the reported complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: révision hip left.The product was not returned to j&j medtech orthopaedics; however photos were provided for review.The photographs attached were reviewed, however they do not represent the reported complaint.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.A manufacturing record evaluation was performed for the finished device product description: dlt ts cer hd 12/14 28mm +5.0 product code: 136528720 lot number: 4350635 and no non-conformances or manufacturing irregularities were identified.The overall complaint was not confirmed as the photographs provided are not representative of the reported complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device product description: dlt ts cer hd 12/14 28mm +5.0 product code: 136528720 lot number: 4350635 and no non-conformances or manufacturing irregularities were identified.Device history review : a manufacturing record evaluation was performed for the finished device product description: dlt ts cer hd 12/14 28mm +5.0 product code: 136528720 lot number: 4350635 and no non-conformances or manufacturing irregularities were identified.
|
| |
|