CARDIAC SCIENCE CORPORATION POWERHEART AED G3 PLUS SEMI AUTOMATIC; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number 9390E-1001
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including daily/weekly/monthly testing, on/off current testing, patient/circuit impedance testing, and shock testing using known good test pads, without duplicating the report.An internal inspection found no discrepancies.The main board will be replaced as a precaution.The device will be recertified and returned to the customer once the repair estimate has been approved.The electrode pads used were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during functional testing, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERHEART AED G3 PLUS SEMI AUTOMATIC
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
CARDIAC SCIENCE CORPORATION
500 burdick parkway
deerfield WI 53531 9692
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key20664256
Report Number2112020-2024-00895
Device Sequence Number20642735
Product Code MKJ
UDI-Device Identifier00812394021123
UDI-Public(01)00812394021123
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number9390E-1001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2024
Initial Report FDA Received Date11/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-