The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including daily/weekly/monthly testing, on/off current testing, patient/circuit impedance testing, and shock testing using known good test pads, without duplicating the report.An internal inspection found no discrepancies.The main board will be replaced as a precaution.The device will be recertified and returned to the customer once the repair estimate has been approved.The electrode pads used were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
|