| Catalog Number |
18965040S |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported: "gamma nail is broken at the entry point of the femoral neck screw.Metal removal necessary." on the x-ray attached in the intake, we see that one of the distal locking screw is also broken.
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: please refer to section d6a implant date.Although initially only a broken nail was reported and confirmed, physical examination revealed that one of the two locking screws was also broken.It was determined that the locking screw with lot code k1bc384 was broken approximately in the center of the shaft.Broken surface presents feature of a fatigue fracture.The intended use of a locking screw is not to take over the intended use of the temporary nail implant in the event of a nail fracture.According to the damage verified and since the nail broke in a fatigue fracture due to axial overload the biomechanical stresses on the whole construct increased and thus, broken locking screw was resulting.For further details and full context please refer to main investigation under complaint mfr # (b)(4).A review of the device history for the broken locking screw lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.According to available information a deficiency of the locking screw was not verified.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
|
| |
|
Event or Problem Description
|
|
As reported: "gamma nail is broken at the entry point of the femoral neck screw.Metal removal necessary." on the x-ray attached in the intake, we see that one of the distal locking screw is also broken.
|
| |
|