| Model Number |
1000186-115 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during preparation of the indeflator device air bubbles were noted; therefore, the indeflator device was not used in the procedure.There was no patient involvement and clinically significant delay reported in the procedure.No additional information was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported leak was unable to be confirmed.Device history record (dhr) and corrective and preventive actions (capa) reviews were performed and revealed an indication of a product quality issue.Additionally, a review of the complaint handling database identified similar incidents from this lot.The investigation determined the reported difficulties appear to be a potential product quality issue.On oct 24th 2024, abbott vascular determined that a field action was required for specific lots of priority pack 20/30 w/115 rhv product.The product associated with this complaint is from the impacted population.This action is being taken due to an increased potential for a leak in the indeflator under pressure or vacuum due to a gap in the hose snap fitting and o-rings from a specific supplier.B3: date of event estimated.
|
| |
|