• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-P180
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a retrospective complaint review involving bf series scopes.This report reflects olympus' re-evaluation of previous medical device reporting decision.The device was returned to olympus for evaluation and the customer's reported issue was confirmed.The plastic distal end cover was found burned/melted around the instrument channel outlet at the distal end.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that high-energy hand instruments (laser/high-frequency/etc) were used without ensuring sufficient distance from the distal end.A definitive root cause cannot be identified.The suggested event can be detected and prevented by handling the device in accordance with ¿chapter 4 operation¿ in the instructions for use.Olympus will continue to monitor the field performance of this device.
 
Event or Problem Description
It was reported the bronchovideoscope's plastic distal end cover was burnt.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20667544
Report Number9610595-2024-23267
Device Sequence Number7305192
Product Code EOQ
UDI-Device Identifier04953170317965
UDI-Public04953170317965
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K061313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-P180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2023
Initial Report FDA Received Date11/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2011-2023 AND Z-0189-2
Patient Sequence Number1
Patient SexUnknown
-
-