• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZEST ANCHORS, LLC LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI); ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZEST ANCHORS, LLC LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI); ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number 02792-SB
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Abutment broke in patients mouth some time after placement.Doctor had to bring patient in to remove the broken screw thread from the implant.Unsure how long after placement the event occurred.Doctor thinks it broke while getting attachment to correct torque.Fractured abutment.
 
Event or Problem Description
Abutment broke in patients mouth some time after placement.Doctor had to bring patient in to remove the broken screw thread from the implant.Unsure how long after placement the event occurred.Doctor thinks it broke while getting attachment to correct torque.Fractured abutment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI)
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
ZEST ANCHORS, LLC
2875 loker ave east
carlsbad
Manufacturer (Section G)
ZEST ANCHORS, LLC
2875 loker ave east
carlsbad
Manufacturer Contact
amie toor
2875 loker ave e
carlsbad 
MDR Report Key20668522
Report Number2023950-2024-00324
Device Sequence Number7239334
Product Code NHA
UDI-Device Identifier00840481109553
UDI-Public00840481109553
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K072878
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number02792-SB
Device Catalogue NumberN/A
Device Lot NumberL31D9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2024
Supplement Date Received by Manufacturer11/01/2024
Initial Report FDA Received Date11/12/2024
Supplement Report FDA Received Date01/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-