E initial reporter cannot be provided due to japanese privacy regulations.Japan medtronic personnel reported event to manufacturer on behalf of the customer.Issue identified during product analysis.H3: device evaluation summary: medtronic conducted an investigation based upon all information received.It was reported that during service this 980 ventilator failed the power on self test (post).The device was available for evaluation.While the service personnel (sp) was performing servicing, confirmed the reported issue and replaced the exhalation valve assembly (eva).The ventilator passed all calibrations and tests per the manufacturer's specifications at the time of service.One eva was returned for failure investigation.A visual inspection was carried out on the returned eva, no anomalies were observed.The part was attached to the test ventilator and powered up.During the calibration, the device failed the pressure sensor cross check test.After a thorough investigation, a faulty autozero solenoid (so1) on the exhalation sensor printed circuit board assembly (pcba) of the eva was identified.Investigation determined if so1 were to fail while in use on a patient, the ventilator response would be to enter backup ventilation (buv).The cause of the reported event was isolated to a faulty autozero solenoid (so1) on exhalation sensor printed circuit board assembly (pcba) of the eva.There is an existing previous internal investigation regarding this issue.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|