| Catalog Number |
205548000 |
| Medical Device Problem Codes |
Material Deformation (2976); Illegible Information (4050)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that trial heads were worn and numbers are illegible.All pieces retrieved, no surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.Corrected: d4 primary udi number.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that trial heads were worn and numbers are illegible.All pieces retrieved, no surgical delay.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the top surface was severely stripped.The investigation did not find enough evidence to confirm the allegation of illegible etch.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not conducted due to not being applicable to the complaint condition.A dimensional inspection was not conducted due to not being applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the s/c & mod cath trl 48/28 would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
|
| |
|
Search Alerts/Recalls
|