• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625315160
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient received discrepant results when testing with his coaguchek xs meter serial number (b)(6) and a second coaguchek xs meter (unknown serial number) used at the doctor's office.This medwatch will apply to the test strips used on the doctor's office meter.Please refer to the medwatch with a1.Patient identifier (b)(6) for information related to the test strips used on the patient's meter.At 10:30 a.M., a sample from the patient was tested using the patient's meter, resulting in a value of 4.4 inr.At 11:30 a.M., a sample from the patient was tested using the doctor's office meter, resulting in a value of 3.1 inr.The patient's therapeutic range is 2.6 - 3.6 inr.
 
Additional Manufacturer Narrative
The patient's meter was requested for investigation and a replacement product was sent to the patient.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Common Device Name
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key20669512
Report Number1823260-2024-03280
Device Sequence Number17951611
Product Code GJS
UDI-Device Identifier00365702128101
UDI-Public00365702128101
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04625315160
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/18/2024
Initial Report FDA Received Date11/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
WARFARIN.
Patient Age79 YR
Patient SexMale
Patient Weight76 KG
-
-