| Model Number |
GIA6038S |
| Medical Device Problem Codes |
Retraction Problem (1536); Component or Accessory Incompatibility (2897)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: gia6048l gia6048l gia 60-4.8sgl use loadingunit - (lot#unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open whipple, during resecting the bowel, after firing the second time, the knife blade did not go down.A new open stapler was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open whipple, during resecting the bowel, the first reload worked fined, the second reload could not be loaded and then after firing the second time with the second reload, the knife blade did not go down.A new open stapler was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted no abnormalities.The sulu (single use loading unit) was fully applied.The interlock was engaged but the knife blade was exposed.Microscopic inspection of the knife blade displayed no damage.The firing knob was not fully retracted.It was reported that the device was difficult to retract the knife blade and it was difficult to load.The reported issue were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.If the firing knob is not fully retracted, difficulty in unloading the loading unit may be experienced, and may prevent further loading from occurring.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: remove the sulu, separate the instrument halves.Holding the cartridge-half of the instrument, grasp the proximal end of the sulu ta bs and pull up and out to remove the sulu from the instrument.To place a new sulu into the instrument, hold the sulu by the proximal end tabs and insert into the cartridge fork at a 30 to 45 degree angle from the distal end down until the unit snaps into place.R emove shipping wedge after sulu is fully loaded.Sulu will ¿float¿ up and down in final loaded position.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|