(b)(4).G2: foreign- japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d2, d9, g3, g6, h1, h2, h3, h6, h11 corrected section: h4 complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified the device has not been cycled and the sutures and anchors remain inside the needle.The needle has fractured.Sem analysis of the juggerstitch straight implant sample showed that it fractured due to possible bending overload.Device is used for treatment.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|