EXACTECH, INC. CC FEMORAL SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
208-01-03 |
| Medical Device Problem Codes |
Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Pain (1994); Synovitis (2094); Osteolysis (2377); Joint Laxity (4526); Implant Pain (4561); Aseptic loosening (4883)
|
| Date of Event |
12/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 14 years and 1 months after a left total knee replacement procedure, the patient has experienced prosthesis wear, osteolysis, synovitis, cyst(s), implant pain and joint laxity.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: e.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial left total knee arthroplasty.Subsequently, it was reported that the patient experienced prosthesis wear, osteolysis, synovitis, cysts, pain, aseptic loosening and joint laxity.As a result, the patient underwent a left knee revision approximately 13 years and 11 months after initial implantation.It was later reported that the patient's femoral component had debonded.No further patient impact reported.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b5, d1, d4, g4, h4, h6 - health effect - clinical code and h11 d10: 208-23-13 - cc tibial insert sz 3, 13mm: (b)(6); 208-09-05 - cc stem ext adaptor 5 degree: (b)(6).Tibial tray.Three peg patella.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|