|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: h4: the device manufacture date is currently unavailable.Investigation summary the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: visual: foreign substance/debris/cleaning/sterilization.Per service reports, this complaint was confirmed.Based on service manual operational and diagnostic, the device was found to be non-repairable, and it was discarded.Based on the investigation findings, the faulty parts were identified as the root cause for the device failure during the service evaluation.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
|
It was reported that postoperatively to an unknown procedure, it was observed that the springs on the coupler ac ffl 22 device had internal corrosion and the clips could not be adjusted.During in-house engineering evaluation, it was determined that device had foreign substance/debris/cleaning/sterilization.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2024.All available information has been disclosed.
|