| Catalog Number |
545035500 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Pain (1994); Synovitis (2094); Swelling/ Edema (4577)
|
| Date of Event |
07/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Doi: (b)(6) 2015: male patient received an attune right tka to treat oa.The patella was resurfaced and depuy cement x2 was utilized.The procedure was completed without complications.Dor: (b)(6) 2021: 67 year-old male patient received a right attune total knee revision to treat pain secondary to loosening of the tibial tray.Upon entering the joint, the surgeon expressed a small effusion and debrided synovitis.The tibial tray was grossly loose and debonded at competitor cement to implant interface and revised.The femoral component was well-fixed but revised.There was no reported product problem with the revised tibial insert.The patella was well-fixed and retained.The patient received a competitor revision tibial construct.The procedure was completed without complications.Doi: (b)(6) 2015.Dor: (b)(6) 2021.Right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank."(b)(4)." if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|