EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA AORTIC BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 3000TFX21MM
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11.Additional narrative.Surgical/percutaneous intervention is indicated or performed, or harm occurred due to the device, or there is a device malfunction that could cause or contribute to a serious injury.This event is considered a serious injury.The device was not returned for evaluation, as its availability is unknown.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Event or Problem Description
It was learned via implant patient registry that a 3000tfx 21mm aortic valve was explanted and replaced with a 3300tfx 27mm after an implant duration of 7 years, 9 months due to unknown reasons.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11.Additional narrative.Updated d4, h4, and h6.Based on the information available, a definitive root cause cannot be conclusively determined.A device history record (dhr) review was not performed, as no information regarding a device failure mode was provided.Furthermore, there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA AORTIC BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
one edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key20674146
Report Number2015691-2024-08586
Device Sequence Number8395244
Product Code DYE
UDI-Device Identifier00690103163643
UDI-Public(01)00690103163643(17)180906
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/06/2018
Device Model Number3000TFX21MM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2024
Supplement Date Received by Manufacturer12/13/2024
Initial Report FDA Received Date11/13/2024
Supplement Report FDA Received Date12/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/07/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-