• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE BLU 30IN 6-0 D/A CC-1 MP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE BLU 30IN 6-0 D/A CC-1 MP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number 8707H
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a vascular procedure on (b)(6) 2024 and suture was used.It was reported a half a box suture has been removed from the shelf as the doctor found that each of the sutures was fraying.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/13/2024.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: how long was the delay? please specify in minutes.Was there any change in the patient¿s post-operative care due to the prolonged procedure? please confirm the quantity of devices involved in the reported event and the quantity of devices available to be returned.When did the reported event occurred (in the package, during removal from the package, during handling prior to using on the patient, during use on the patient or post-op)? please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep).Please provide the status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details.The following information was reported: was surgery delayed due to the reported event? yes.If yes, number of minutes: unknown.Action taken when event occurred? new suture different lot.Was procedure successfully completed? yes.Were fragments generated? unknown.If yes, were they removed easily without additional intervention? unknown.Patient status/ outcome / consequences: no.Was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown.Is the patient part of a clinical study: unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/21/2024 the following information was received: only 10 sterile sutures from this box returned versus half the box reported.Returning the 10 sutures for analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 12/10/2024.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with ten representative unopened samples was received for analysis.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to the customer complaint and no defects were observed during evaluation.Also, a tactile test was performed with the affected parts of the thumb and forefinger and no anomalies were detected.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 30IN 6-0 D/A CC-1 MP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20674556
Report Number2210968-2024-11946
Device Sequence Number21070112
Product Code GAW
UDI-Device Identifier10705031019461
UDI-Public(01)10705031019461(17)280131(10)TBBHKX
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8707H
Device Lot NumberTBBHKX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2024
Supplement Date Received by Manufacturer11/14/2024
12/10/2024
Initial Report FDA Received Date11/13/2024
Supplement Report FDA Received Date11/21/2024
12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-