SYNTHES GMBH GUIDE BLOCK FOR 2 COLUMN PLATE 6 HOLE HEAD/LEFT; GUIDE, SURGICAL, INSTRUMENT
|
Back to Search Results |
|
| Catalog Number |
03.111.601 |
| Medical Device Problem Codes |
Break (1069); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, the patient underwent an unknown surgery.In the surgery, when a nurse attempted to connect the guiding block to the plate, the plate and the guiding block did not engage no matter how many times the nurse tried.When the surgeon tried, the block connection broke off with the threading engaged in the plate.Since the threaded portion of the guiding block remained on the plate in question, a plate one size smaller (left small 3-hole left ex small 3-hole) was used for fixation without any problem.The surgery was completed successfully within 30 minutes surgical delay, patient status/ outcome reported as stable.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the guidingblock f/2-column dist-rad-pl2.4 6 was broken from the shaft of the screw.The broken fragment remines stuck inside of the plate hole.Therefore, it was not possible to evaluate the condition of treads.The shaft of the screw was slightly deformed.In addition, the head of the screw had signs of normal use.No other issue was identified.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined based on the evidence provided.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the guidingblock f/2-column dist-rad-pl2.4 6 would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part:03.111.601.Lot:200229-401.Manufacturing site: werk selzach.Supplier:(b)(4).Release to warehouse date:12 jan 2021.Expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|